Skip to content

Public FDA record

Z-0664-2012: Bivona Uncuffed Pediatric Flex Tend Plus Extra Length Tracheostomy Tube Sizes 4.0 - 6.0 (With a V-Shaped

Recall record Z-0664-2012 for Smiths Medical ASD, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0664-2012
Company in FDA record
Smiths Medical ASD, Inc.
FDA classification date
Jan 19, 2012
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Difficulty disconnecting accessories from the Bivona Uncuffed Neonatal, Pediatric and Flextend Tracheostomy Tubes TR3 connector, which may result in a delay in initiating therapy or patient discomfort. If excessive force is used to detach the accessory, this may result in decannulation of the tube which could result in serious injury or death; especially if no replacement tracheostomy tube is imm

Source facts

Action
Smith Medical sent an Urgent Field Safety Notice dated November 28, 2011, via U.S. Mail to all affected consignees. The notice identified the product, description of problem, and actions to be taken. This notice included a Customer Information Bulletin and Urgent Field Safety Notice Confirmation Form. Customers were instructed to 1) identify all affected unused product in inventory and segregate it to a quarantine location; 2) complete and return the attached Confirmation Form by fax to 219-989-7259 or email to bivona.tr3@smiths-medical.com; and 3) distributors were instructed to stop distributing and provide Smith Medical with a list of their customers, including contact details, so that Smith Medical can notify the customers of this notice. For questions contact Smiths Medical Customer Service at 800-258-5361, Option 1.
Initiated
2011-11-28
Root cause
Device Design
Terminated
2014-10-17
Distribution
Worldwide Distribution - USA Nationwide: AZ, CA, CO, CT,DE, FL, GA, HI, IL, IN, KS, KY, LA, SC, TX, OR, NJ, MO, UT, MI, NY, NM, MA, MN, MS, NC, ND, NE, NH, NV, OH, OK, OR, RI, SD, TN, VA, VT, WA, WI, WV including Washinton DC, Australia, Austria, Barbados, Belgium, Brazil, Canada, Chile, Cyprus, Czech Republic, Denmark, Egypt, Finland, France, Germany, Gibraltar, Great Britain (UK), Greece, Hong Kong, Ireland, Israel, Italy, Kuwait, Libya, Malaysia, Netherlands, New Zeland, Norway, Poland, Portugal, Qatar, Russian Federation, Saudi Arabia, Singapore, Spain, Sweden, Switzerland, United Arab Emirates, United States, Vatican City State
Recall number
Z-0664-2012
Recall status
Terminated
CDRH record ID
106114
FDA classified
2012-01-19
Recall event number
60449

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 21, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 21, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 21, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗