Affected lots were manufactured with the protective release liner in an incorrect orientation. When attempting to remove this disposable release liner, a clear film may be left on the grounding pad which may interfere with patient skin contact and lead to uneven heating.
Source facts
Action
The firm initiated the recall by letter hand delivered by an Abbott field representative. The letter requested that consignee cease use of the affected lot and return the product with the aid of a field representative.
Initiated
2019-10-14
Root cause
Process control
Terminated
2021-07-19
Distribution
nationwide
Recall number
Z-0664-2020
Recall status
Terminated
CDRH record ID
177902
FDA record created
2019-12-10
Recall event number
84272