Shots were terminated by the mAs integrator; however, the 4 ms exposure time did not allow enough tolerance to achieve diagnostic exposures due to variations in hardware and how different timers are started.
Source facts
Action
The firm will notify consignees using a service bulletin that was already posted. Additionally, the firm will implement a firmware update to extend the timer (4 ms to 10 ms minimum) to ensure even short exposures achieve the necessary image quality for diagnostic use.
Initiated
2023-05-31
Root cause
Software design
Distribution
Worldwide and US Nationwide Distribution: NJ, PR, CA
Recall number
Z-0664-2025
Recall status
Open, Classified
CDRH record ID
210755
FDA classified
2024-12-12
Recall event number
95603