The internal speaker components are failing and do not produce an audio alarm in the event of patient emergency.
Source facts
Action
Customers of record were notified of the advisory by letter sent via US Mail on October 18, 2004, via certified mail. A second letter was sent on December 9, 2004, advising customers that upgrades would be available.
Upgrades would be available via returns to the firm''s service center, upgrade kits and speaker replacements via self-installation, and a third party service provider.
Initiated
2004-10-18
Root cause
Other
Terminated
2006-05-15
Distribution
There are 2,950 Direct US and US territory Customers. Product shave also been distributed to Canada, Europe, Japan, Latin America and Australia.
Recall number
Z-0665-05
Recall status
Terminated
CDRH record ID
37556
FDA classified
2005-03-30
Recall event number
31271