Skip to content

Public FDA record

Z-0665-2020: Healix Knotless ADV BR 4.75 Suture Anchor

Recall record Z-0665-2020 for DePuy Mitek, Inc., a Johnson & Johnson Co.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0665-2020
Company in FDA record
DePuy Mitek, Inc., a Johnson & Johnson Co.
FDA record creation date
Dec 10, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Specific product codes and lots of the Mitek HEALIX ADVANCE" Knotless Anchors were manufactured with BIOCRYL" Biocomposite Material instead of BIOCRYL RAPIDE Biocomposite Material. BIOCRYL and BIOCRYL RAPIDE are biocompatible and effective material choices for absorbable, injection-molded sports medicine implant fixation.

Source facts

Action
On November 4, 2019 the firm distributed Urgent Medical Device Recall Communication letters by mail: URGENT FIELD SAFETY NOTICE PRODUCT RECALL HEALIX ADVANCE" Knotless Anchors Dear Valued Customer, On 04-November-2019, DePuy Mitek, Inc. initiated a Voluntary Product Removal of specific product codes and lots of the Mitek HEALIX ADVANCE" Knotless Anchors. The anchors were manufactured with BIOCRYL" Biocomposite Material instead of BIOCRYL RAPIDE Biocomposite Material. BIOCRYL and BIOCRYL RAPIDE are biocompatible and effective material choices for absorbable, injection-molded sports medicine implant fixation. No complaints or adverse events related to this issue have been received for the affected product codes/lots. Product Code: 222330 Product Description: HEALIX KNOTLESS ADV BR 4.75 GTIN: 10886705023462 Lot Number: 5L45257 Product Code: 222331 Product Description: HEALIX KNOTLESS ADV BR 5.5 GTIN: 10886705023479 Lot Numbers: 5L13543 and 5L45259 Potential Harms: Based on the analysis, it is expected that sport medicine implants made from BIOCRYL (PLLA+TCP) would perform equal to those made from BIOCRYL RAPIDE (PLGA + TCP) during the surgery and therapeutic healing period. As such, this product issue presents no foreseeable increased risk to patient safety nor does it alter the product benefit-risk profile. BIOCRYL and BIOCRYL RAPIDE Comparison: " Each material is used clinically today, but in different DePuy Mitek implant brands; both materials have a more than 15 years of use in clinical applications. BIOCRYL was first launched in 2002 and BIOCRYL RAPIDE in 2004. " The mechanical properties (strength, stiffness) of BIOCRYL and BIOCRYL RAPIDE are similar at the time of anchor deployment; BIOCRYL offers equivalent in vitro strength retention during the therapeutic healing period and out to 12 weeks. " As for long-term absorption rates, BIOCRYL RAPIDE is expected to absorb in approximately two years versus approximately four years
Initiated
2019-11-04
Root cause
Employee error
Terminated
2020-05-05
Distribution
US: AL AZ CA CT FL IA IL IN LA MA MD ME MI MN MT NC NE NH NJ NM OH OK OR PA TN TX VA WA WI International: Belgium, Chile, China, Czech Republic, France, India, Japan, Korea, Latvia, New Zealand, Slovakia Sweden
Recall number
Z-0665-2020
Recall status
Terminated
CDRH record ID
177815
FDA record created
2019-12-10
Recall event number
84237

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗