The pumps will stop, show an overspeed error message, and will not restart using routine procedures if the user reduces the pump speed to zero RPM while the pump is operating in pulse mode.
Source facts
Action
Recall letters dated 3/17/05 were sent to customers advising them of the problem, the temporary fix, and that software will be upgraded.
Initiated
2005-03-17
Root cause
Other
Terminated
2005-09-16
Distribution
Nationwide, Australia, Belgium, Canada, Egypt, Germany, Hong Kong, India, Indonesia, Japan, Malasia, Mexico, Pakistan, South Africa, South Korea, Taiwan, Thailand and United Arab Emirates.
Recall number
Z-0666-05
Recall status
Terminated
CDRH record ID
37978
FDA classified
2005-03-31
Recall event number
31489