Z-0666-2025: MICRO-X Rover Mobile X-ray System, # MXU-RV19

Recall record Z-0666-2025 for Micro-X Ltd.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0666-2025
Company in FDA record
Micro-X Ltd.
FDA classification date
Dec 12, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

During internal testing it was found that if the system is Ready to expose and the hand switch is then dropped into the stowage unit such that it lands on the prep/expose switch, the prep switch can be activated for a very short time such that the orange ring of death is displayed in the tube cooling area of the screen. The system then becomes unresponsive to kV/mAs changes, prep/expose presses. The only way to proceed is to suspend the patient, Exit Nexus DR and relogin and continue the exam.

Source facts

Action
Starting on 10/23/24 Service Bulletin SB002-22 was sent to customers informing them that a field upgrade was released for the installation of an exposure switch holder. Customers were asked to complete and return an acknowledgement response form to support-us@micro-x.com For further information, contact your local service representative at support-us@micro-x.com, 1-206-249-8764.
Initiated
2024-10-23
Root cause
Device Design
Distribution
US Nationwide Distribution: TX, CA, OH, ID, WA, AZ
Recall number
Z-0666-2025
Recall status
Open, Classified
CDRH record ID
210715
FDA classified
2024-12-12
Recall event number
95593

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗