Incorrect result: The absorbance has changed as the standards have aged; the standards' turbidity has become denser. The recalled lot is now reading 0.15 ABS @ 625 nm. TREK'S internal specification and the manufacturer's (Hardy Diagnostics) specification is 0.08 - 0.10 ABS @625 nm. CLSI specifies 0.08 - 0.13 ABS @ 625 nm.
Source facts
Action
On 9/22/2008 TREK Diagnostic sent a recall notification letter to their customers advising them of the problem with the product. They instructed their customers to inventory their stock for any remaining recall lot and immediately destroy.
Initiated
2008-09-22
Root cause
Nonconforming Material/Component
Terminated
2010-07-06
Distribution
USA distribution in OH and IL.
Recall number
Z-0669-2009
Recall status
Terminated
CDRH record ID
74791
FDA classified
2009-01-28
Recall event number
50129