Open original FDA record ↗Recall record
- FDA identifier
- Z-0669-2017
- Company in FDA record
- Terumo Medical Corp
- FDA record creation date
- Nov 29, 2016
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
During recent routine testing, the MISAGO Peripheral Stent was found not to conform to product specification related to stent diameter at the proximal/distal ends and/or the shape of the stent.
Source facts
- Action
- Terumo Medical Corporation issued an Urgent Product Recall Bulletin dated August 4, 2016, via Fed Ex. Customers must immediately discontinue use and quarantine any available stock of the Misago peripheral stent to prevent further use. The customers should complete the enclosed Inventory Confirmation form as soon as possible and email or fax the completed form to misago.rc@terumomedical.com or (734) 330-2034, Attn: MISAGO Recall. Customers are encourage you to contact Terumo Medical Corporation with any questions or concerns -Terumo Medical Corporation Quality Department: 1.800.283.7866, Press 4 for list of departments, Press 7 for Quality Assurance; Monday Friday, 8 a.m. 5 p.m. ET - Email Quality Assurance: misago.rc@terumomedical.com - Customer Care: 1.800.888.3786 - Contact your local Terumo representative
- Initiated
- 2016-08-04
- Root cause
- Device Design
- Terminated
- 2017-10-19
- Distribution
- Nationwide
- Recall number
- Z-0669-2017
- Recall status
- Terminated
- CDRH record ID
- 150185
- FDA record created
- 2016-11-29
- Recall event number
- 75370