- FDA identifier
- Z-0669-2018
- Company in FDA record
- Synthes, Inc.
- FDA record creation date
- Feb 21, 2018
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
May have difficulty attaching the drill sleeve to the plate. If the drill sleeve will not connect to the plate intraoperatively, surgical delay could occur. If the pilot hole is drilled without the drill sleeve, the pilot hole could be off-center which could result in screw and/or plate loosening.
Source facts
- Action
- In August of 2017 customers were mailed an URGENT FIELD SAFETY NOTICE titled "884115 - Locking Compression Plate (LCP) Drill Sleeve" requesting that customers who received product subject to recall to please contact: RA-DPYCH-flactionint@ITS.JNJ.com or call 610-719-5450. In the event that you have product requiring use: Pre-assemble the drill sleeve with the plate preoperatively. If you experience difficulty connecting the drill sleeve to the plate, use the additional 1.1 mm locking drill sleeve provided in the graphic case, or: o Align the 1.1 mm drill bit to the plate as perpendicular as possible and then drill freehand and the locking screw can be engaged in this manner. o If the drill hole must be used and no option to lock remains, a non-locking screw can be used.
- Initiated
- 2017-08-07
- Root cause
- Process design
- Terminated
- 2018-07-25
- Distribution
- Distribution to US, Canada, and Germany.
- Recall number
- Z-0669-2018
- Recall status
- Terminated
- CDRH record ID
- 161199
- FDA record created
- 2018-02-21
- Recall event number
- 79021