Skip to content

Public FDA record

Z-0669-2020: Elecsys Anti-HAV IgM assay-for the in vitro qualitative detection of IgM antibodies to hepatitis A virus

Recall record Z-0669-2020 for Roche Diagnostics Operations, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0669-2020
Company in FDA record
Roche Diagnostics Operations, Inc.
FDA record creation date
Dec 12, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Current software version of the Anti-HAV IgM APP-file (version 06.05-101) downloaded an erroneous Border Area setting, (1.0 1.00) instead of (0.90 1.10).

Source facts

Action
Roche notified consignees via Urgent Medical Device Correction TP-00811 mailed via UPS Ground (receipt required) on 10/28/19. Letter states reason for recall, health risk and action to take: Consignee Instructions: 1. Follow the instructions provided in the Actions to be Taken by Customers/Users Currently Using APP-file Version 06.05-101 section of this communication. 2. Confirm your cobas e 801 module is using the updated APP-file xx.xx-102 (version 102). If it is not, immediately update to APP-file xx.xx-102 (version 102). 3. If your facility has distributed the affected product to another site, please ensure this Urgent Medical Device Correction (UMDC) is provided to that site. 4. Complete the attached fax form and fax or email it according to the instructions on the form. 5. File this Urgent Medical Device Correction for future use. The UMDC was also posted to the diagnostics.roche.com website. Questions contact the Roche Support Network Customer Support Center, 24 hours a day, seven days a week at 1-800-428-2336.
Initiated
2019-10-28
Root cause
Employee error
Terminated
2022-06-28
Distribution
US Nationwide distribution.
Recall number
Z-0669-2020
Recall status
Terminated
CDRH record ID
177850
FDA record created
2019-12-12
Recall event number
84260

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗