Open original FDA record ↗Recall record
- FDA identifier
- Z-0669-2025
- Company in FDA record
- Hologic, Inc
- FDA classification date
- Dec 11, 2024
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
IFU was not provided with the Sizer Set sterilization tray Revised Sterilization Instructions for Sizer Set
Source facts
- Action
- Hologic issued Letter-Revised Sterilization Instructions for Sizer Set on 10/10/22 via Fed'X. The letter states reason for recall, health risk and action to take: Effective immediately, please steam autoclave F0101 in accordance with such IFU, subject to the following parameters. Any additional parameters contained in Hologic s IFU for F0101 should no longer be utilized. link to their Instructions for Use (IFU) below: https://summitmedicalusa.com/wordpress/wp-content/uploads/2021/05/40192-01_RevZ_IS-IFU_Multilingual.pdf Please also note that when sterilizing F0101, one (1) layer of wrapping material (or one layer of 1 ply wrap) is necessary to provide a sterile barrier for each sterilization tray. Please do not utilize 2 layers or one layer of 2-ply wrap. In the future, Hologic will be updating the IFU supplied with the sizers within F0101, to reflect this change. mailto:BreastHealth.Support@hologic.com
- Initiated
- 2022-10-10
- Root cause
- Labeling Change Control
- Distribution
- US Nationwide distribution.
- Recall number
- Z-0669-2025
- Recall status
- Completed
- CDRH record ID
- 211201
- FDA classified
- 2024-12-11
- Recall event number
- 95781