Monaco is using the incorrect energy when optimizing and calculating dose.
Source facts
Action
An Important Field Safety Notification letter is being emailed to affected customers beginning 11/22/2019. The letter identifies affected product, provides a description of the problem, and recommends user not use multiple energies within the same plan until a software fix has been applied to the device. A response form was asked to be returned. Affected customers are advised to contact local Elekta service teams for questions.
Initiated
2019-11-22
Root cause
Software Design Change
Distribution
IA, IN, NY, OH, TX, WI.
Internationally distributed to: Australia, Canada, China, Denmark, Germany, Greece, Hong Kong, Italy, Netherlands, Sweden, Switzerland, United Kingdom.
Recall number
Z-0670-2020
Recall status
Open, Classified
CDRH record ID
178080
FDA record created
2019-12-12
Recall event number
84330