LIFEPAK CR Plus defibrillators with software version 1.16 or lower may miscalculate the HLC battery capacity and result in the CHARGE-PAK battery charger icon/indicator to display and falsely indicate battery needs replacement.
Source facts
Action
Firm sent letters on 03/02/2007 dated 03/2007 via certified return receipt to consignees with affected LIFEPAK CR Plus devices. The firm advised consignees to upgrade their devices' software, and indicated that each consignee will be provided a user-installable software upgrade for their device(s) within 45 days.
Initiated
2007-03-02
Root cause
Software design
Terminated
2012-04-26
Distribution
Distribution is worldwide. The device is designed for non-professional users and are primarily used in public access defibrillator environments such as airports, schools, or public libraries.
Recall number
Z-0671-2007
Recall status
Terminated
CDRH record ID
50629
FDA classified
2007-03-29
Recall event number
37490