Skip to content

Public FDA record

Z-0671-2025: BD Pyxis MedStation ES 7 Drawer Auxiliary Tower, REF: 324

Recall record Z-0671-2025 for CareFusion 303, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0671-2025
Company in FDA record
CareFusion 303, Inc.
FDA classification date
Dec 11, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to a software issue, automated dispensing cabinet devices may open the wrong cubie pocket or position, which can lead to: 1) Unintended and incorrect medication removal, or 2) Potential for a different medication (medication pulled in error) to be refilled/loaded, which may lead to may lead to low inventory or a stockout.

Source facts

Action
On 11/14/24, correction notices were mailed/emailed to customers: Devices can continue to be used in accordance with the following: Best practice is for providers (e.g., nurses) to verify the medication being removed is accurate against the hospital patient record (source of truth) prior to administration1. o Labeling includes the following warning: Always utilize the BD Pyxis ES System patient/order information in conjunction with the medical record when making clinical decisions. Making clinical decisions solely from System information may result in errors in care resulting in serious injury or death. " To further aid that verification step, it has been estimated that BCMA (Barcode Medication Administration) can reduce medication administration errors. Similarly, "Scan on Remove" functionality can be enabled within the MedStation to help capture an erroneously dispensed medication. " Safe storage and labeling for neuromuscular blocking (NMB) agents - Institute for Safe Medication Practices (ISMP) recommends to segregate, sequester, and differentiate NMBs from other medications. This can help identify a high-risk medication is being removed in error. " If a different medication is dispensed, the error may be caught during the check process prior to leaving the pharmacy. If error is not caught, pharmacy technician will still refill/load the medication dispensed properly at the device. " ES server includes notifications for those medication pockets that are empty within the System to trigger pharmacy to refill. Complete and return response form to BDRC1@bd.com For further assistance, contact firm: Technical Support - 1-800-727-6102, 24 hours, 7 days a week; Regional Complaint Center, 1-844-823-5433, Mon-Fri 8:00am-5:00pm CT, productcomplaints@bd.com; Post Market Quality Operations, Fax: 1-410-316-4258, Email: BDRC1@bd.com On 2/6/2025, updated correction notices were sent notifying customers that the firmware update was available.
Initiated
2024-11-14
Root cause
Software design
Distribution
Worldwide - US Nationwide distribution including in the states of OK, LA, FL, AR, VA, KY, PA, AL, NC, CA, MN, AK, TX, MI, ME, OH, OR, NY, IL, RI, GA, WI, NH, KS, CO, WA, MO, MD, AZ, SC, NV, WV, NM, IN, NJ, TN, DE, MT, MA, VT, DC, NE, CT, ID, WY, MS, UT, HI, IA, SD, ND, GU, MP, PR and the countries of CA, BR, JP, ES, NZ, AU, FR, AR, MX, IT, CL, TR, SA, MY, CH, IE, TH, SG, KW, BE, DE, PT, QA, MA, FI, AE, EG, BS, GB, BM, GR, BH, KR, MC, PL, LB, HK, IN, IL, MO, IQ, TN, NO, PH, CO, TW, CN, JO, OM.
Recall number
Z-0671-2025
Recall status
Open, Classified
CDRH record ID
211005
FDA classified
2024-12-11
Recall event number
95643

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.