An error in the Bio-Tek ELx800 microplate reader programming specifications for the ENA screen Assay exists in that an equivocal (borderline) zone is missing.
Source facts
Action
Customers were sent a notification letter informing customers of an error in the Bio-Tek ELx800 programming specifications. Customers were asked to repeat testing for historical runs for RSLT (S/CO) that read from 0.95 to 1.0. A response form is asked to be faxed back to DiaSorin.
Initiated
2004-04-20
Root cause
Other
Terminated
2006-07-22
Distribution
CT, FL, NY, WA, TX, AR, TN, CANADA
Recall number
Z-0672-05
Recall status
Terminated
CDRH record ID
36809
FDA classified
2005-04-06
Recall event number
30920