Z-0672-2024: SPRYSTEP VECTOR KAFO, Custom made device dynamic ankle-foot orthosis (KAFO) non locking, REF #s: 17VKCUS

Recall record Z-0672-2024 for Townsend Design. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0672-2024
Company in FDA record
Townsend Design
FDA classification date
Jan 5, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to potential air pockets between layers of material, their is the potential for complete struct breakage of knee-ankle-foot orthosis.

Source facts

Action
On November 7, 2023, Townsend Design (Thuasne USA), issued a "Urgent Medical Device Recall notification to affected consignees: Townsend Design asked consignees to take the following actions: 1. Clinician should contact any patient who was fit with a device received between June 14th and September 10, 2023 to notify them of the recall and schedule a visit to examine the patient's device. 2. During the patient visit, the clinician should examine the device for structural defects (delamination, cracking, complete breakage, etc.) in the area identified in the picture above in Section 3 3. If the device is observed with a defect, patient should immediately discontinue use of the device and it should be returned to Townsend Design (Thuasne USA) 4. Clinician can contact Townsend Design (Thuasne USA) Customer Service at 800-432-3466 for return and replacement instructions 5. Townsend Design (Thuasne USA) will either refund or replace the product 6. Clinician to complete and return the attached Acknowledgement and Receipt Form within 5 working days
Initiated
2023-09-23
Root cause
Process control
Distribution
US: AL AR AZ CA CO CT DC DE FL GA IA ID IL IN KENT KS KY LA MA MD ME MI MN MO MS MT NC NE NH NJ NM NV NY OH OK OR PA QC SC SD TN TX UT VA VT WA WI WV WY, Puerto Rico, OUS: None OUS: Canada, New Zealand, Japan, Australia, Netherlands, United Kingdom
Recall number
Z-0672-2024
Recall status
Completed
CDRH record ID
203862
FDA classified
2024-01-05
Recall event number
93299

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗