The data reduction program written for a specific automated microtiter plate instrument (Bio-Tek ELx800 reader) was found to be faulty. This program is used to calculate the cut off values for Diastat Anti-Mitochondrial Antibody Kit. Program should include a borderline specification to indicate repeat testing is needed.
Source facts
Action
A notification letter and effectiveness check form was sent to consignees. This letter recommended that users manually verify results using the specifications outlined in the package insert until a new software version is installed.
Initiated
2004-04-12
Root cause
Other
Terminated
2006-07-22
Distribution
CA, CT, KS, MA & WA
Recall number
Z-0673-05
Recall status
Terminated
CDRH record ID
36810
FDA classified
2005-04-06
Recall event number
30922