Z-0673-2025: Access Intact PTH assay on the DxI 9000 Access Immunoassay Analyzer, Catalog number A16972, for the quant

Recall record Z-0673-2025 for Beckman Coulter Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0673-2025
Company in FDA record
Beckman Coulter Inc.
FDA classification date
Dec 11, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The reason for the recall is that EDTA and lithium heparin plasma samples that are stored in the DxI 9000 Access Immunoassay Analyzer sample wheel for a prolonged period before performing the intact PTH (parathyroid hormone) test may produce falsely decreased intact parathyroid hormone results. This could lead to a missed diagnosis, a delay in treatment, or inappropriate clinical management, depending upon the patient s clinical condition.

Source facts

Action
On November 8, 2024, the firm notified affected customers via URGENT MEDICAL DEVICE RECALL letters that the affected assay may produce falsely decreased results if EDTA and lithium heparin plasma samples are stored in the DxI 9000 Access Immunoassay Analyzer sample wheel for a prolonged period before performing the test. Access Intact PTH assay Routine Mode and Intraoperative Mode users can select any of the following options: . Use serum as the sample type. . Perform testing on an alternate methodology, such as the Access 2 or UniCel DxI 600 or 800 Immunoassay Analyzers. . Continue to use EDTA and/or lithium heparin plasma sample types if the tests are completed within the timeframes listed below upon sample presentation to the DxI 9000 Access Immunoassay Analyzer (see page 3 for instructions): o Access Intact PTH assay Routine Mode: 25 minutes. o Access Intact PTH assay Intraoperative Mode: 20 minutes. . Beckman Coulter recommends sharing the content of this letter with your laboratory and/or medical director regarding the need to perform a retrospective review of the patient results generated from Access Intact PTH Routine Mode plasma samples on the DxI 9000 to determine if retesting is necessary.
Initiated
2024-11-08
Root cause
Process design
Distribution
Foreign distribution to Albania, Australia, Austria, Bahrain, Belgium, Brazil, C¿te d'Ivoire, Croatia, Czechia, Egypt, France, Germany, Ghana, Hungary, India, Ireland, Israel, Italy, Kazakhstan, Kuwait, Malaysia, Netherlands, New Zealand, Poland, Portugal, Slovakia, South Africa, S. Korea, Spain, Switzerland, Taiwan, United Kingdom of Great Britain and Northern Ireland.
Recall number
Z-0673-2025
Recall status
Open, Classified
CDRH record ID
211080
FDA classified
2024-12-11
Recall event number
95738

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗