Data reduction Program written for specific sutomated microtiter plate insturment (Bio-Tek ELx800 reader) is faulty in that the positive result limit is to be greater than 15 U/ml, rather than greater than or equal to 15 U/ml.
Source facts
Action
Customers were sent notification letters and effectiveness check form. recommended users manually veify positve results (RSLT), using product insert specifications.
Initiated
2004-10-26
Root cause
Other
Terminated
2006-07-22
Distribution
CT, MI, MT, UT & CANADA
Recall number
Z-0674-05
Recall status
Terminated
CDRH record ID
36812
FDA classified
2005-04-06
Recall event number
30924