Markings reversed: Graduated markings on telescoping bone distractors were reversed, causing the distractors to compress rather than distract the treatment site when adjusted.
Source facts
Action
Orthofix notified sales agents by phone starting on 09/28/07. Sales agents were to return product for replacements.
Initiated
2007-09-28
Root cause
Process control
Terminated
2008-03-20
Distribution
Nationwide Distribution - USA states: CA, FL, GA, IL, PA, MI, NY and WI.
Recall number
Z-0674-2008
Recall status
Terminated
CDRH record ID
66393
FDA classified
2008-02-26
Recall event number
45878