Skip to content

Public FDA record

Z-0675-2024: MYNXGRIP 6F/7F, REF MX6721, Vascular Closure Device (VCD). Used to seal femoral arterial and femoral veno

Recall record Z-0675-2024 for Cordis US Corp. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0675-2024
Company in FDA record
Cordis US Corp
FDA classification date
Jan 5, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Cordis became aware that the label for the carton box containing 10 units is incorrect for some boxes of two lots of MynxGrip. The carton box from lot F2322902 (Item MX5021, MynxGrip 5F) as labeled as lot F2322903 (MX6721, MynxGrip 6/7F) and product from lot F2322903 (MX6721, MynxGrip 6/7F) was labeled as F2322902 (Item MX5021, MynxGrip 5F). The primary label on individual devices is correct.

Source facts

Action
Cordis issued an Urgent MEDICAL DEVICE RECALL notice to its consignees on 11/28/2023 via via FEDEX. The notice explained the issue and requested the return of any affected product in inventory. For questions related to the recall and/or acknowledgement form that are not adequately addressed in the letter, please contact Cordis QA at: GMB-CordisFieldAction@cordis.com or call Cordis Customer Support 1-800-327-7714 Option 2, Monday through Friday from 8:00 AM to 8:00 PM EDT.
Initiated
2023-11-28
Root cause
Labeling Change Control
Distribution
US: NC, WV, TN, MS, MI, OH, CA
Recall number
Z-0675-2024
Recall status
Open, Classified
CDRH record ID
204974
FDA classified
2024-01-05
Recall event number
93617

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗