Skip to content

Public FDA record

Z-0676-06: Gemcal Reference Serum (magnesium calibrator). Calibrator, Multianalyte Mixture. Level 1 and Level 2 Ch

Recall record Z-0676-06 for Alfa Wassermann, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0676-06
Company in FDA record
Alfa Wassermann, Inc
FDA classification date
Mar 23, 2006
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Inaccurate test results. The magnesium values for the A31/A41 Gemcal Serum calibrator gave control results that were biased approx. 20% high.

Source facts

Action
Field correction letters were sent to the firm's customers via FAX on 1/14/2006-1/17/2006 and via UPS on 1/17/2006-1/19/2006. An updated letter was sent out via FAX on 1/27,30,31/2006 and via UPS on 1/30/2006 and 2/1/2006.
Initiated
2006-01-26
Root cause
Other
Terminated
2006-04-18
Distribution
The kits were sent nationwide. The product was shipped to Alfa Wasserman's sister firm in Wergen, Netherlands. There are several US Navy accounts.
Recall number
Z-0676-06
Recall status
Terminated
CDRH record ID
44474
FDA classified
2006-03-23
Recall event number
34687

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗