Inaccurate test results.
The magnesium values for the A31/A41 Gemcal Serum calibrator gave control results that were biased approx. 20% high.
Source facts
Action
Field correction letters were sent to the firm's customers via FAX on 1/14/2006-1/17/2006 and via UPS on 1/17/2006-1/19/2006. An updated letter was sent out via FAX on 1/27,30,31/2006 and via UPS on 1/30/2006 and 2/1/2006.
Initiated
2006-01-26
Root cause
Other
Terminated
2006-04-18
Distribution
The kits were sent nationwide. The product was shipped to Alfa Wasserman's sister firm in Wergen, Netherlands. There are several US Navy accounts.
Recall number
Z-0676-06
Recall status
Terminated
CDRH record ID
44474
FDA classified
2006-03-23
Recall event number
34687