Skip to content

Public FDA record

Z-0676-2008: Philips Medical Systems -- Cardiac Viewer or Pulmonary Viewer software application version 2.0 in use wit

Recall record Z-0676-2008 for Phillips Nuclear Medicine. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0676-2008
Company in FDA record
Phillips Nuclear Medicine
FDA classification date
Feb 8, 2008
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Incorrect Measurement: The firm discovered a software defect in Cardiac and Pulmonary Viewer software application options used in the Precedence CT Sub-System. The defect causes incorrect measurements in derived images which are zoomed and saved in a batch file.

Source facts

Action
A Customer Advisory Notification was mailed to affected consignees on September 28, 2007. The firm informed consignees that incorrect measurement can occur when the Cardiac Viewer or Pulmonary Viewer is used to zoom and save derived images in batch files. The firm provided instructions for operators to prevent incorrect measurements. Phillips is currently investigating methods to correct this issue and will provide updates, free of charge, when they become available. If you have questions concerning this recall, please contact Phillips Medical at 1-800-722-9377, (option5: Diagnostic Imaging, option 5:NM).
Initiated
2007-09-28
Root cause
Software design
Terminated
2009-02-04
Distribution
Worldwide Distribution --- USA including states of FL, NY, VT, MD, TX, and MO and country of Canada.
Recall number
Z-0676-2008
Recall status
Terminated
CDRH record ID
65274
FDA classified
2008-02-08
Recall event number
45329

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗