Open original FDA record ↗Recall record
- FDA identifier
- Z-0676-2008
- Company in FDA record
- Phillips Nuclear Medicine
- FDA classification date
- Feb 8, 2008
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Incorrect Measurement: The firm discovered a software defect in Cardiac and Pulmonary Viewer software application options used in the Precedence CT Sub-System. The defect causes incorrect measurements in derived images which are zoomed and saved in a batch file.
Source facts
- Action
- A Customer Advisory Notification was mailed to affected consignees on September 28, 2007. The firm informed consignees that incorrect measurement can occur when the Cardiac Viewer or Pulmonary Viewer is used to zoom and save derived images in batch files. The firm provided instructions for operators to prevent incorrect measurements. Phillips is currently investigating methods to correct this issue and will provide updates, free of charge, when they become available. If you have questions concerning this recall, please contact Phillips Medical at 1-800-722-9377, (option5: Diagnostic Imaging, option 5:NM).
- Initiated
- 2007-09-28
- Root cause
- Software design
- Terminated
- 2009-02-04
- Distribution
- Worldwide Distribution --- USA including states of FL, NY, VT, MD, TX, and MO and country of Canada.
- Recall number
- Z-0676-2008
- Recall status
- Terminated
- CDRH record ID
- 65274
- FDA classified
- 2008-02-08
- Recall event number
- 45329