Z-0676-2020: MRIdian Linac Radiation Therapy System Model 10000 and 20000

Recall record Z-0676-2020 for Viewray, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0676-2020
Company in FDA record
Viewray, Inc.
FDA classification date
Dec 12, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Issue with transferring treatment regiment from one system to another. Adequate instructions for transferring treatment plan(s) are not provided within the operator's manual or within the software operating system.

Source facts

Action
On 03/03/2019, the firm emailed an "URGENT DEVICE RECALL" to customers to inform them of a safety issue with the Treatment Planning and Delivery System (TPDS) in the MRIdian System. The notice explained that when a partially-completed patient treatment regiment is moved to their system from another system, the use needs to rescale the plan. The TPDS does not automatically rescale the treatment plan. The customers are instructed: If they want to move a treatment plan from one system to another and still show the total original prescribed dose on the MRIdian TPDS the customer will have multiple ways to do this. 1. Once the user/operator has evaluated the plan at the full dose, ensure the prescription matches the dose and number of fractions to be delivered on the MRIdian, and scale the dose to the target by either renormalizing or changing the target objective in the optimizer. 2. Keep the original total number of fractions (including those delivered elsewhere) in the prescription but when the plan is assigned to the fractions in the delivery calendar, Discontinue the number of fractions that have already been delivered on another system. 3. Import the previously delivered dose files as a RTDose and fuse with the plan images. Optimize to the remaining target dose that is to be deliver on MRIdian. Include the imported dose as Previously Delivered Dose and Calculate in the Finish screen to display the full dose which has and will be delivered. Ensure the prescription matches the dose and number of fractions to be delivered on the MRIdian. 4. Import the previously delivered dose files as RTDose and fuse with the plan images. Optimize to the original total dose, including the Previously Delivered Dose as part of the optimization. Ensure the prescription matches the total dose and number of fractions including those treated elsewhere and those to be delivered on the MRIdian. The customer is also informed that the Recalling Firm will address this is
Initiated
2019-03-03
Root cause
Error in labeling
Terminated
2021-04-28
Distribution
US - CA, FL, IL, MI, MO, NY, and WI OUS: Denmark, France, Germany, Italy, Japan, Netherlands, Turkey, South Korea, and UAE
Recall number
Z-0676-2020
Recall status
Terminated
CDRH record ID
177991
FDA classified
2019-12-12
Recall event number
84303

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗