Skip to content

Public FDA record

Z-0676-2024: BD Recanalization System, REF: BDRECANSYSTEM

Recall record Z-0676-2024 for Bard Peripheral Vascular Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0676-2024
Company in FDA record
Bard Peripheral Vascular Inc
FDA classification date
Jan 5, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Recanalization systems may be in a manufacturing mode that allows the system to continuously run without the typical 30 second pauses, and 4-minute warning/5-minute restart required messages to indicate the ten-minute maximum catheter use time has been exceeded, which may increase the risk of the device overheating or breaking.

Source facts

Action
On 11/14/23 recall notices were mailed and emailed to customers who were asked to do the following: 1) Discontinue use of affected product and immediately quarantine until ready to return for service activities. 2) Share this notification with all device users within your facility and network to ensure they are aware of this field action. This notification should also be shared with any organization where the potentially affected devices have been transferred. 3) Complete and return the response form via email to BDRC49@bd.com 4) Recalling firm will contact customers to coordinate product return for service. Customers with questions can call the firm's North American Regional Complaint Center at 1-844-823-5433, Say "Recall" when prompted M-F 8am - 5pm CT, or email productcomplaints@bd.com
Initiated
2023-11-14
Root cause
Process design
Distribution
US: CA, OH, PA, NY, TX, NJ, MS, IL, GA, AL, AZ, NC, NV, PR, IN, MA, WI, KS, AR, FL, MD, IA, NE, LA, HI, MO, DE, DC, VA, NM, CO, KY, ME, MN
Recall number
Z-0676-2024
Recall status
Open, Classified
CDRH record ID
204958
FDA classified
2024-01-05
Recall event number
93607

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.