Z-0677-2024: BrightView, Gamma Camera System, Product Code 882480.

Recall record Z-0677-2024 for Philips North America. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0677-2024
Company in FDA record
Philips North America
FDA classification date
Jan 5, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

While using Pre-Programmed Motion during an extrinsic quality assurance scan, a gap is created between the patient support and the detector, which may present a potential extremity entrapment hazard. Risk to patients may include fracture, body part loss of function/debilitation, muscle or ligament sprain or strain, laceration, crush injury, abrasion, or contusion.

Source facts

Action
On November 30, 2023, the firm notified customers of the issue through Urgent Medical Device Correction Letters. Users were instructed to monitor the patient during system movement when Pre-Programed Motion is in use and review the warning and caution guidance information provided with the recall notice, and which is also outlined in the Instructions for Use. Philips will contact you to schedule a time for a field service engineer to visit your site and install a technical solution to resolve this issue. If you need any further information or support concerning this issue, please contact your local Philips representative: For North America, contact the Customer Care Solutions Center (1-800-722-9377).
Initiated
2023-11-30
Root cause
Device Design
Distribution
Domestic distribution nationwide. Foreign distribution worldwide.
Recall number
Z-0677-2024
Recall status
Open, Classified
CDRH record ID
205007
FDA classified
2024-01-05
Recall event number
93643

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗