Z-0677-2025: DiaSorin LIAISON Q.S.E.T. Device Plus, REF 319060; intended for use in the preparation of human stool spe

Recall record Z-0677-2025 for Diasorin Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0677-2025
Company in FDA record
Diasorin Inc.
FDA classification date
Dec 12, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Customer complaints were received alleging leaking devices and insufficient buffer volume. Diasorin's internal investigation determined that 0.14% of the LIAISON Q.S.E.T. Device Plus in the lots identified have loose clear caps, which may allow buffer to leak from the tube. A stool sample prepared with less than the correct amount of buffer may result an incorrectly high patient result.

Source facts

Action
DiaSorn issued an Urgent Field Safety Notice - Medical Device Recall notice to its consignees on 11/07/2024 via email. The notice explained the issue with the product, risk to health, and requested the following: "Ac xons to be taken by the Customer/User " No need to review past patient results. " Devices should be inspected for loose clear caps prior to use. The user should confirm that the clear cap is tight by grasping the cap and twisting. If the cap is loose, the device should be discarded. Do not attempt to use the device as the loose clear cap may have allowed buffer to leak, causing insufficient buffer volume. " Customers should forward this communication to required individuals within the organization or to any organization where the potentially affected devices have been distributed. " Diasorin will replace any tubes found with loose clear caps. Contact Diasorin Product Support for replacement product." For questions or concerns, contact: Diasorin Product Support: 1-800-328-1482 Select Option 1 (U.S. and Canada); 1-651-351-5888 (Fax); productsupport@diasorin.com
Initiated
2024-11-07
Root cause
Process control
Distribution
Worldwide
Recall number
Z-0677-2025
Recall status
Open, Classified
CDRH record ID
211310
FDA classified
2024-12-12
Recall event number
95829

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗