Z-0678-2020: Maquet Cardiopulmonary (MCP)/Getinge ROTAFLOW Drive, Model Number 701022161. The device is a component o

Recall record Z-0678-2020 for Maquet Cardiovascular Us Sales, Llc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0678-2020
Company in FDA record
Maquet Cardiovascular Us Sales, Llc
FDA record creation date
Dec 12, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

A loose ODU plug / Coaxial cable connection may result in fluctuating flow values on the Console.

Source facts

Action
An Urgent Medical Device Recall notification letter dated 12/3/19 was sent to customers. If your affected RFD unit is in clinical use, switch to RPM mode. Using hospital patient management protocols and physician judgement, perform the following steps: 1. Check your inventory for an unaffected RFD unit . If an unaffected RFD unit is available, it should be at the clinicians discretion to transfer the patient from the affected unit to the unaffected unit. 2. If an unaffected RFD unit is unavailable or if temporary interruption of extracorporeal blood flow is not possible check the following options: a. Please check your inventory for an independent external flow measurement system to measure the actual flow. Only use a reliable external flow measurement that is gauged and zeroed. Use that measurement system in conjunction with the RPM mode. If necessary adjust the RPM according to the flow reading as obtained with the independent external flow measurement system. A continuous control of the patient vital signs monitoring is indicated. If the the Bubble Sensor intervention is already active, leave the setting in place. b. In case no external flow measurement system is available stay on RPM mode and continuously control vital signs monitoring. It is indicated to closely monitor the blood gases of the respective patient. The assessment of the vital sign monitoring and the blood gas analysis in conjunction with the indicated pump speed (RPM) is required. If the the Bubble Sensor intervention is already active, leave the setting in place. 3. When treatment of patient is completed, remove your affected RFD unit from use and follow the Actions to be Taken for return and repair below. If your affected RFD unit is not in clinical use, remove your affected RFD unit and follow the Actions to be Taken for return and repair below. Actions to be taken 1. Our records indicate that you have received a Rotaflow Drive Unit with a serial number that is affected by this
Initiated
2019-12-03
Root cause
Nonconforming Material/Component
Terminated
2020-09-18
Distribution
The products were distributed to the following US states: ME, NV, NY, OH, PA, TX, and VA.
Recall number
Z-0678-2020
Recall status
Terminated
CDRH record ID
178194
FDA record created
2019-12-12
Recall event number
84377

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗