Z-0678-2025: Hybrid Recon software on Hermes Medical Imaging Suite version 6.1 Hybrid Recon software versions affected

Recall record Z-0678-2025 for Hermes Medical Solutions AB. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0678-2025
Company in FDA record
Hermes Medical Solutions AB
FDA classification date
Dec 12, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due a potential software/configuration issue that may result is incorrect alignment during reconstructing a SPECT/CT study.

Source facts

Action
On 10/31/2024, the firm emailed a "Field Safety Notice (FSN)" letter to customers informing them that when reconstructing a SPECT/CT study, in some cases, the SPECT and CT series are not correctly aligned. This may be due to separately performed acquisitions, or due to the camera operator re-setting the reference point (landmark) between SPECT and CT acquisitions. Normally, the Hybrid Recon application warns the user if the Frames of Reference do not match, however in some configurations this warning message may be turned off. Customers are instructed how to avoid the problem by: In all cases, the user must verify correct SPECT to CT alignment by careful inspection of the fused images during reconstruction. To ensure that you are warned about possible SPECT to CT misalignment, the Skip Frame of Reference warning radio button shall not be set on . The Skip Frame of Reference warning radio button can be found in the Program Parameters window (button at top right corner of the window left side of the ? symbol). Go to the Recon tab and find the Skip frame of reference warning radio button. If the button is on as shown here, this warning will not be shown to the users. The setting of the Skip Frame of Reference warning radio button should be checked to ensure it is off for each workflow Oncology, Neurology, Cardiology and Lung. Please contact your Hermes Medical Solutions support if you do not have permission to make the changes required. For questions or further assistance contact General: support@hermesmedical.com Canada: support.ca@hermesmedical.com USA: support.us@hermesmedical.com
Initiated
2024-10-31
Root cause
Device Design
Distribution
U.S.: CA, CT, FL, HI, KY, MA, MD, MO, NJ. NY, OH, TN, TX, VA, and WA O.U.S.: Germany, and United Kingdoms
Recall number
Z-0678-2025
Recall status
Open, Classified
CDRH record ID
210923
FDA classified
2024-12-12
Recall event number
95673

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗