Open original FDA record ↗Recall record
- FDA identifier
- Z-0679-05
- Company in FDA record
- The Spineology Group
- FDA classification date
- Apr 6, 2005
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Three malfunctions, due to jamming of bone material in the fill tube occurred, which allowed the fill tube to advance more deeply than intended into the surgical site. Fill tubes are used in delivery of bone graft materials.
Source facts
- Action
- A Product Update, January 2005 (Vol. 2) was e-mailed to all representatives and also sent by mail on January 28, 2005. Telephone calls to each distributor/representative were made by Spineology Sales & Marketing staff between January 28 and January 31. The Product Update informs customers that they will be receiving new tubes and new platens soon and that any tubes and platens currently in their possession should be returned to Spineology.
- Initiated
- 2005-01-28
- Root cause
- Other
- Terminated
- 2005-11-17
- Distribution
- 11 consignees located in: Arkansas, Arizona, California, Georgia, Illinois, Michigan, Minnesota, North Carolina, New York, Oklahoma and United Kingdom
- Recall number
- Z-0679-05
- Recall status
- Terminated
- CDRH record ID
- 37373
- FDA classified
- 2005-04-06
- Recall event number
- 31198