Skip to content

Public FDA record

Z-0679-2020: Roche Homocysteine-In vitro test for the quantitative determination of total L-homocysteine in human seru

Recall record Z-0679-2020 for Roche Diagnostics Operations, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0679-2020
Company in FDA record
Roche Diagnostics Operations, Inc.
FDA record creation date
Dec 13, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Homocysteine Reagent, Calibration Failures and Quality Control Recovery Issues on cobas c 701 and 702 modules

Source facts

Action
The Urgent Medical Device Correction (UMDC) was sent via Roche issued Urgent Medical Device notification via FedEx (signature required) to the consignee list on 10/04/2019. The letter states issue, health risk nd action to take: Discontinue use and discard the remaining affected product in inventory according to local waste guidelines. Because a replacement lot is currently not available, Roche will issue credit for Homocysteine cobas c packs, lot number 422313, remaining in your inventory. If customers have another validated system (i.e. cobas c501/502) in their lab and wish to set up the Roche Homocysteine assay on one of these modules (different catalog number than that subject to this recall), they will be instructed to contact a Roche Field Specialist for support. " A reimbursement for send out testing to an accredited laboratory will also be provided for those customers without another Roche system available. Contact the Roche Support Network Customer Support Center, 24 hours a day, seven days a week at 1-800-428-2336 if you have questions about the information The UMDC will also be posted to the diagnostics.roche.com website.
Initiated
2019-10-04
Root cause
Device Design
Terminated
2020-10-09
Distribution
US Nationwide distribution in the states of NJ, HI, IA.
Recall number
Z-0679-2020
Recall status
Terminated
CDRH record ID
177008
FDA record created
2019-12-13
Recall event number
84018

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗