Z-0680-06: Baxter BM11A Blood Monitor Pump hemodialysis machine; Baxter Healthcare Corporation, Deerfield, IL 60015;

Recall record Z-0680-06 for Baxter Healthcare Renal Div. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0680-06
Company in FDA record
Baxter Healthcare Renal Div
FDA classification date
Mar 29, 2006
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Advisory to alert users to assure that the hemofilter of the blood line is in the horizontal position to avoid kinking of the bloodline tubing to avoid hemolysis, and to show the compatible blood tubing sets to be used with the BM11/BM11A Blood Monitor Pump and BM25 System ultrafiltration hemodialysis system.

Source facts

Action
Important Product Information letters dated 12/21/05 were sent to all BM-series instrument users, informing them of the potential for kinking of the tubing if the hemofilter is in the vertical position and the compatibility testing conducted on certain blood tubing sets for use with the BM-series instruments and Baxter''s recommendation that the following sets be used with the BM-series instruments: a) product code 5M3006M for use with the BM 11 and the BM 11a, when the BM 11a is used as a stand alone device; b) product code 5M3048 for use with the BM 11a when operated as part of the BM 25 System; and c) product code 5M3046 for use with the BM 14 part of the BM 25 System. Please note that the BM 14 cannot be used as a stand alone instrument. A fourth bloodline manufactured by Nextron and identified as a BM11 Low Volume Tubing (Baxter product code 5M3132 and Nextron code DHZ0006) was found to be unqualified for use for the BM series instruments. A recommended blood tubing sets and setup guide with diagrams was included with the letter. Both the letter and the setup guide contained the following warning: ''During machine set up, verify that the selected blood tubing set has been recommended for use with the Baxter bm11/bm11a Blood Monitor Pump and the Baxter bm25 System. Using non-recommended blood tubing sets, even if the blood tubing set fits and appears to function correctly, may have clinical consequences. These consequences may include blood tubing kinking, which may cause hemolysis. Clinical consequences of returning hemolyzed blood to a patient range from back pain and shortness of breath up to and including death.'' The users were also informed that the warning was being included in a new revision of the Operator''s Manual, and were requested to call Baxter''s Center for Service at 1-888-229-0001 to return any stocks of product code 5M3132 for credit. Any questions were directed to the Center for One Baxter at 1-800-422-9837.
Initiated
2005-12-21
Root cause
Other
Terminated
2008-06-23
Distribution
Nationwide including Puerto Rico, and internationally to Colombia, Hong Kong and China
Recall number
Z-0680-06
Recall status
Terminated
CDRH record ID
43806
FDA classified
2006-03-29
Recall event number
34245

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗