Skip to content

Public FDA record

Z-0680-2017: Olympus Scientific Solutions Americas Corporation (OSSA) Model: VANTA Analytical X-Ray System

Recall record Z-0680-2017 for Olympus Scientific Solutions Americas. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0680-2017
Company in FDA record
Olympus Scientific Solutions Americas
FDA record creation date
Dec 7, 2016
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Possible unintentional x-ray emission after users attempt early termination of the sequence.

Source facts

Action
The Center for Devices and Radiological Health (CDRH) understands that the CAP involving 41 units located in the United States will consist of the following: 1. Shipments were stopped until a software correction was designed and validated. The software correction was designed by October 27, 2016, and was implemented on November 7, 2016 for all units that had not left the place of manufacture and for all new products. 2. The software correction will be made available to all distributors, sales personnel and users. Issue 1 is corrected by altering the behavior of the stop button and trigger such that they stop the current test and terminate the test sequence. To correct issue 2 the software will be modified to grey-out the Stop button during this very small period making it consistent with rest of the VANTA Start/Stop test UI paradigm. Both during user Start and Stop actions, users are shown a greyed button to acknowledge user Click and the button is un-greyed when the system completes the user operation. Both user operations typically take 100 400 milliseconds to complete. 3. All VANTA units in the United States are currently in the possession of Olympus or your distributors. No units are in the possession of users. 4. Olympus or its distributors will apply the software correction prior to delivering a VANTA system to any purchasers in the United States. For further questions, please call (781) 419-3900.
Initiated
2016-10-27
Root cause
Radiation Control for Health and Safety Act
Terminated
2020-07-30
Distribution
US Distribution to the states of : TX, WA, GA, MA, SC, AZ, PA, VA, CA, OH, NJ, MD, CO, IL, LA, and MN.
Recall number
Z-0680-2017
Recall status
Terminated
CDRH record ID
150838
FDA record created
2016-12-07
Recall event number
75575

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗