Skip to content

Public FDA record

Z-0680-2020: VisuMax Software Version 2.10.13 with activated Software-Module ReLEx (SMILE) option used in conjunction

Recall record Z-0680-2020 for Carl Zeiss Meditec, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0680-2020
Company in FDA record
Carl Zeiss Meditec, Inc.
FDA record creation date
Dec 13, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

VisuMax devices with software version 2.10.13 and activated SW-Module ReLEx (SMILE) option may experience a software issue. In case of a suction loss during treatment the software allows the user to choose the option immediate restart or restart treatment. Suction loss can occur in phase 1 through phase 5. The software defect refers to Phase 2 (between 10% and 100% of lower lenticule cut) only. The software offers a flap cut, but due to the software issue it performs a cap cut instead of a flap cut if the user proceeds.

Source facts

Action
Consignees were notified with an Urgent Medical Device Recall letter dated 12/06/2019. The letter identified affected devices, described the software issue, and provided direction on what steps to take with suction loss in phase 2, until a software update is available. A response form was asked to be returned. Questions can be directed to 877-486-7473.
Initiated
2018-12-06
Root cause
Software design
Terminated
2021-02-02
Distribution
US nationwide distribution.
Recall number
Z-0680-2020
Recall status
Terminated
CDRH record ID
176793
FDA record created
2019-12-13
Recall event number
83913

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗