Open original FDA record ↗Recall record
- FDA identifier
- Z-0680-2020
- Company in FDA record
- Carl Zeiss Meditec, Inc.
- FDA record creation date
- Dec 13, 2019
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
VisuMax devices with software version 2.10.13 and activated SW-Module ReLEx (SMILE) option may experience a software issue. In case of a suction loss during treatment the software allows the user to choose the option immediate restart or restart treatment. Suction loss can occur in phase 1 through phase 5. The software defect refers to Phase 2 (between 10% and 100% of lower lenticule cut) only. The software offers a flap cut, but due to the software issue it performs a cap cut instead of a flap cut if the user proceeds.
Source facts
- Action
- Consignees were notified with an Urgent Medical Device Recall letter dated 12/06/2019. The letter identified affected devices, described the software issue, and provided direction on what steps to take with suction loss in phase 2, until a software update is available. A response form was asked to be returned. Questions can be directed to 877-486-7473.
- Initiated
- 2018-12-06
- Root cause
- Software design
- Terminated
- 2021-02-02
- Distribution
- US nationwide distribution.
- Recall number
- Z-0680-2020
- Recall status
- Terminated
- CDRH record ID
- 176793
- FDA record created
- 2019-12-13
- Recall event number
- 83913