Open original FDA record ↗Recall record
- FDA identifier
- Z-0680-2025
- Company in FDA record
- Spectranetics Corporation
- FDA classification date
- Dec 16, 2024
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Due to complaints received for the marker band detaching from the distal tip of the catheter and remaining in the patient as laser catheter was retracted.
Source facts
- Action
- On 11/25/2024, the firm sent an "URGENT MEDICAL DEVICE CORRECTION" Letter via USPS certified mail to customer informing them that Philips has become aware of a potential safety issue with a limited number of Turbo-Elite Laser Atherectomy Catheters where the marker band has the potential to become detached from the device. Detached marker band are visible under fluoroscopy and should remain on the guidewire. Since a detached marker band requires additional intervention to retrieve, temporary or medically reversible adverse health consequences are possible. The potential clinical harms of a detached marker band are embolization, dissection, or perforation, which are harms listed in the Adverse Event Procedural Complications section of the Instructions for Use. Actions to be taken by Customer: Philips recommends the continued use of the Turbo-Elite Laser Atherectomy Catheter and to follow the Instructions for Use, including using fluoroscopy to ensure the marker band is not left behind after retracting the device. As a precaution, if a marker band detaches and remains on the wire in the patient, then additional intervention to retrieve the marker band is required. There are multiple possible endovascular interventions to retrieve a detached marker band including snaring, balloon catheter inflation, and using a filter wire to capture the marker band. If these techniques are unsuccessful, then stenting can be used to secure the marker band in place to avoid distal embolization. Ensure you report all occurrences of this issue to Philips via a complaint. For questions or assistance-contact local Philips representative or Philips Image Guided Therapy Devices Customer Service at 1-800-231-0978, option 2.
- Initiated
- 2024-11-25
- Root cause
- Device Design
- Distribution
- U.S.: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, ILIN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD TN, TX, UT, VA, WA, WI, and WV. O.U.S.: Australia, Belgium, Brazil, China, Croatia, Cyprus, Czech Republic, Denmark, France, Germany, Italy, Japan, Kuwait, Netherlands, Poland, Spain, Sweden, Switzerland, United Arab Emirates, and United Kingdom.
- Recall number
- Z-0680-2025
- Recall status
- Open, Classified
- CDRH record ID
- 211283
- FDA classified
- 2024-12-16
- Recall event number
- 95815