Skip to content

Public FDA record

Z-0681-2024: Alinity c Processing Module REF 03R67-01 The Alinity c Processing Module is a fully automated chemistry

Recall record Z-0681-2024 for Abbott Laboratories. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0681-2024
Company in FDA record
Abbott Laboratories
FDA classification date
Jan 9, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The devices contain a dry natural rubber (latex) subcomponent, which are not labeled as containing dry natural rubber(latex).

Source facts

Action
On 10/19/2023, the firm sent an "Urgent Medical Device Correction" Letter via FedEx priority overnight express to customers informing them that Abbott has determined that, A labeling update is required for the products listed above to meet US regulatory requirement 21 CFR 801.437 (User labeling for devices that contain natural rubber) and Brazilian requirement RDC 37/2015 (medical devices and IVDs to include in the label a Standardized Latex Warning in Portuguese CONTAIN NATURAL LATEX. MAY CAUSE ALLERGIES ). These Abbott products include internal hardware parts and/or subassemblies that contain dry natural rubber (latex) which may cause allergic reactions, but do not include the required precautionary label. Additionally, the Alinity s System Solid Waste Container Rubber Band (LN 04U97-01), used to secure the biohazard bag to the solid waste container, contains dry natural rubber (latex). Customer are instructed to: 1) Follow good laboratory practice by wearing impervious gloves and other personal protective equipment when working on the above listed Abbott product(s). 2) If you have forwarded the product(s) listed above to other laboratories, please inform them of this Product Correction and provide a copy of this letter. 3) Please retain this letter for your laboratory records. A precautionary label will be applied to the affected products. An Abbott representative will schedule a service visit when the instrument label is available for application. For questions or concerns, contact Customer Service at 1-877-4ABBOTT (24 hr/7 days a week).
Initiated
2023-10-19
Root cause
Under Investigation by firm
Distribution
U.S.: AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, Puerto Rico, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, and WV O.U.S.: Brazil
Recall number
Z-0681-2024
Recall status
Open, Classified
CDRH record ID
204144
FDA classified
2024-01-09
Recall event number
93355

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗