Skip to content

Public FDA record

Z-0682-2012: STAT 2 Pumpette I.V. Compensating Controller, Catalog/REF No. PS2, ASSEMBLED IN MEXICO, CONMED CORPORATIO

Recall record Z-0682-2012 for ConMed Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0682-2012
Company in FDA record
ConMed Corporation
FDA classification date
Jan 11, 2012
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

ConMed received 16 complaints for STAT 2 Pumpette devices for inaccurate flow rates and for leakage from the body of the compensating controller. The investigation of the complaints determined that the STAT 2 Pumpette Compensating Controller on several complaint samples had been assembled incorrectly.

Source facts

Action
ConMed sent an "Urgent Medical Device Recall" letters and Reply Forms dated April 15, 2011 to the domestic consignees via UPS Priority Mail pouches and to the foreign consignees via FedEx International Priority. The letter identified the product, description of problem, and actions to be taken. Consignees were instructed to review their inventory and immediately stop use of the affected devices and return them to ConMed Corporation. Consignees who had distributed the affected products outside of their facility were instructed to immediately notify their customers of the recall and provide them with a copy of this notice. Consignees were asked to complete and return the enclosed Reply Form. For questions call 315-624-3237 or email ivcontroller@conmed.com
Initiated
2011-04-15
Root cause
Device Design
Terminated
2012-11-20
Distribution
Worldwide Distribution - USA (nationwide) - including the countries of: Australia, Barbados, Belgium, Bermuda, Canada, Egypt, Israel, India, Italy, Korea, New Zealand, Philippines and Saudi Arabia.
Recall number
Z-0682-2012
Recall status
Terminated
CDRH record ID
106252
FDA classified
2012-01-11
Recall event number
60703

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 21, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 21, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗