Skip to content

Public FDA record

Z-0682-2025: StatStrip Glucose Ketone Hospital Meter System. Model/Catalog Number: 63683. Prescription use blood gluco

Recall record Z-0682-2025 for Nova Biomedical Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0682-2025
Company in FDA record
Nova Biomedical Corporation
FDA classification date
Dec 20, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

A software bug within the meter firmware leads to transmission of erroneous glucose and/or ketone patient test results to a healthcare institution's data management system (DMS) at healthcare sites using WiFi if a user navigates to the Review Results screen while the meter is in the process of transmitting the current test result to the DMS.

Source facts

Action
Nova Biomedical notified consignees via letter on 11/20/2024. Consignees were instructed that all current Nova StatStrip customers that have gone Live with their new meters recently received a meter software update (v.0.0.13.45 or above) from Nova Biomedical to eliminate the potential risk described above. All remaining customers that have been shipped StatStrip meters impacted by this field correction, will have their meter software updated by Nova Biomedical or local dealer prior to going "Live" with their new meters.
Initiated
2024-11-20
Root cause
Software design
Distribution
Distribution US Nationwide including Puerto Rico, Germany, Italy, Israel, Switzerland, Australia, Great Britain, and the Netherlands.
Recall number
Z-0682-2025
Recall status
Open, Classified
CDRH record ID
211233
FDA classified
2024-12-20
Recall event number
95783

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗