Skip to content

Public FDA record

Z-0683-2016: IBA Dosimetry GmbH Lift Table Type 4320, Catalog HA03-010#2 and HA03-000#2

Recall record Z-0683-2016 for Iba Dosimetry Gmbh. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0683-2016
Company in FDA record
Iba Dosimetry Gmbh
FDA classification date
Jan 21, 2016
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

When the electric lift table is connected to the mains power, in a very rare occasion, the table may start to move up in an uncontrolled manner.

Source facts

Action
The field safety notice including a confirmation of receipt was sent to the affected hospitals; by email on 12/10/2015 then followed by 12/17/2015 sent by Fed Ex. The firm requested the mains power be disconnected after setting up the table and BP. Replacement part and instructions will follow.
Initiated
2015-12-10
Root cause
Component design/selection
Terminated
2018-08-08
Distribution
Nationwide and International: Algeria, Angola, Arab Republic of Egypt, Austria, Belarus, Belgium, Bulgaria, Bosnia and Herzegovina, Croatia, Czech Republic, Denmark, Ethiopia, France, Germany, Ghana, Greece, Hungary, Iraq, Ireland, Islamic Republic Iran, Italy, Kazakhstan, Kenya, Kingdom of Saudi Arabia, Kuwait, Lebanon, Libya, Morocco
Recall number
Z-0683-2016
Recall status
Terminated
CDRH record ID
142869
FDA classified
2016-01-21
Recall event number
73036

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗