- Recall classification
- Class II
Reason stated in the source
When the electric lift table is connected to the mains power, in a very rare occasion, the table may start to move up in an uncontrolled manner.
Source facts
- Action
- The field safety notice including a confirmation of receipt was sent to the affected hospitals; by email on 12/10/2015 then followed by 12/17/2015 sent by Fed Ex. The firm requested the mains power be disconnected after setting up the table and BP. Replacement part and instructions will follow.
- Initiated
- 2015-12-10
- Root cause
- Component design/selection
- Terminated
- 2018-08-08
- Distribution
- Nationwide and International: Algeria, Angola, Arab Republic of Egypt, Austria, Belarus, Belgium, Bulgaria, Bosnia and Herzegovina, Croatia, Czech Republic, Denmark, Ethiopia, France, Germany, Ghana, Greece, Hungary, Iraq, Ireland, Islamic Republic Iran, Italy, Kazakhstan, Kenya, Kingdom of Saudi Arabia, Kuwait, Lebanon, Libya, Morocco
- Recall number
- Z-0683-2016
- Recall status
- Terminated
- CDRH record ID
- 142869
- FDA classified
- 2016-01-21
- Recall event number
- 73036