Skip to content

Public FDA record

Z-0684-2020: Walgreens At-Home A1C Test Kit, Software Version Revision D, REF Number 3056

Recall record Z-0684-2020 for Polymer Technology Systems. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0684-2020
Company in FDA record
Polymer Technology Systems
FDA record creation date
Dec 13, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

A numerical value less than 4% or greater than 13% may be displayed instead of the expected <4% or >13% result, due to a software bug.

Source facts

Action
A Customer Bulletin dated November 2016 was distributed to customers.
Initiated
2016-12-02
Root cause
Software design
Terminated
2020-07-16
Distribution
The products were distributed to the following US states: AR, CA, FL, IL, IN, KY, MN, NC, NJ, NY, OH, PA, RI TN, UT, VA, WA, and Puerto Rico. The products were distributed to the following foreign countries: Australia, Bahamas, Belgium, Bahrain, Bangladesh, Canada, China, Columbia, Czech Republic, England, Egypt, Greece, Spain, India, Indonesia, Haiti, Hong Kong, Italy, Japan, Jordan, Kenya, Lebanon, Mexico, Myanmar, Malaysia, Philippines, Pakistan, South Africa, Nigeria, Poland, Saudi Arabia, Sweden, Turkey, Singapore, Trinidad & Tobago, Slovenia, Vietnam, United Arab Emirates, Uruguay.
Recall number
Z-0684-2020
Recall status
Terminated
CDRH record ID
177804
FDA record created
2019-12-13
Recall event number
84231

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗