Open original FDA record ↗Recall record
- FDA identifier
- Z-0685-2013
- Company in FDA record
- DePuy Spine, Inc.
- FDA classification date
- Jan 16, 2013
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
A component within the handle of the Expedium Offset Torque Wrench is prematurely wearing, potentially resulting in the instrument recording torque values outside of the specification.
Source facts
- Action
- DePuy Spine issued an Urgent - Voluntary Product Recall Notification dated July 24, 2012, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to disconstinue the use of the affected product and to remove the EXPEDIUM Offset Torque Wrenches from service and return to DePuy Synthes Spine until direction can be provided by DePuy Synthes Spine regarding a replacement torque limiting device. For questions regarding this recall call 508-828-3769.
- Initiated
- 2012-07-24
- Root cause
- Device Design
- Terminated
- 2013-12-30
- Distribution
- Worldwide Distribution - USA (nationwide) and internationally to Belgium, Egypt, Ireland, Italy, Portugal, and the UK.
- Recall number
- Z-0685-2013
- Recall status
- Terminated
- CDRH record ID
- 111486
- FDA classified
- 2013-01-16
- Recall event number
- 62719