Open original FDA record ↗Recall record
- FDA identifier
- Z-0686-2019
- Company in FDA record
- Elekta, Inc.
- FDA record creation date
- Dec 27, 2018
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
If a Setup field (CT, kV, MV or MVCT) is created by COPYING a treatment field or another setup field, then the copied field will have the same isocenter as the parent and that isocenter may not be the correct isocenter for imaging when it is sent to the machine.
Source facts
- Action
- Elekta notified customers on about 12/04/2018 via letter titled, "Urgent Important Field Safety Notification." The letter contains recommended user action and that appropriate personnel working with the device should be advised on the content of the letter and the letter should be placed in an area accessible to all users. The letter stated that the issue will be resolved in Service Packs to MOSAIQ Releases 2.64, 2.65 and 2.70. Customers will be informed when the fix is available through a Product Bulletin. An acknowledgement Form was sent as part of the letter to be completed and returned no later than 30 days. The firm can be contacted at 408-830-8000 or 855-693-5358.
- Initiated
- 2018-12-04
- Root cause
- Software design
- Terminated
- 2026-02-23
- Distribution
- Distributed in the United States and outside of the Unites States: Algeria; Antigua and Barbuda; Argentina; Australia; Belgium; Botswana; Brazil; Canada; China; Colombia; Denmark; France; Germany; India; Indonesia; Iran; Ireland; Israel; Italy; Japan; Latvia; Lebanon; Malaysia; Mexico; Morocco; Myanmar; Netherlands; New Zealand; Norway; Peru; Philippines; Poland; Portugal; Russian Federation; Rwanda; Saudi Arabia; Senegal; Singapore; Slovenia; South Africa; Spain; Sweden; Switzerland; Taiwan; Thailand; Turkey; United Kingdom; Viet Nam
- Recall number
- Z-0686-2019
- Recall status
- Terminated
- CDRH record ID
- 169582
- FDA record created
- 2018-12-27
- Recall event number
- 81705