Z-0687-2018: Nitinol TC Reusable Electrodes (TCN), Model Numbers: 1)TCN-10, 2) TCNK-10-R*, 3)TCN-10-3M, 4) TCNK-15, 5)

Recall record Z-0687-2018 for Cosman Medical, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0687-2018
Company in FDA record
Cosman Medical, LLC
FDA record creation date
Feb 22, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

After multiple reprocessing cycles, the epoxy resin which holds the TCN Electrode in the hub can exhibit signs of damage. In some cases, this damage may result in the inability to fully remove blood and/or tissue residuals prior to cleaning and re-sterilizing the device.

Source facts

Action
The firm, Cosman Medical, sent an "Urgent Medical Device Correction" notice and Field Correction Reply Verification Tracking Form dated July 5, 2017 to their customers. The notice described the product, problem and actions to be taken. The Customers were instructed to do the following: 1. Immediately discontinue use of devices that have already been used. Used TCN Electrodes should be disposed of in accordance with your institutions infectious material/biohazard waste control procedures. 2. Unused/un-opened TCN Electrodes should be identified as single use only (for one patient). These devices are supplied as non-sterile which require cleaning and sterilization prior to use. These TCN Electrodes may be used once the cleaning and sterilization steps are completed. After single use, the used TCN Electrode should be disposed of in accordance with your institutions infectious material/biohazard waste control procedures. 3. Complete and return the Field Correction Reply Verification Tracking Form. via Email: Cosmanorders@bsci.com or via Fax to: 781-272-6563 - Identify your email with TCN Electrodes Action in the subject line. Customers with questions may call 1-888-8COSMAN or email Cosmanorders@bsci.com.
Initiated
2017-07-05
Root cause
Device Design
Terminated
2018-09-14
Distribution
Worldwide Distribution: US (nationwide) and country of Canada.
Recall number
Z-0687-2018
Recall status
Terminated
CDRH record ID
160150
FDA record created
2018-02-22
Recall event number
78704

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗