False Results- The reagent of the LDL-Cholesterol assay shows a carryover effect on the Lipase assay. Results may be falsely elevated by up to 50 units per liter when used on COBAS INTEGRA 400 or COBAS INTEGRA 400+ analyzers.
Source facts
Action
Consignees were sent and Urgent Medical Device Correction notice dated 12/03/07. The notice provided instructions for conducting an additional wash cycle whenever both the LDL-Cholesterol assays and Lipases assay are used on the COBAS INTEGRA 400 or COBAS INTEGRA 400+ analyzers. Contact Roche Technical Support at 1-800-428-2336 for questions regarding these instructions.
Initiated
2007-12-03
Root cause
Process design
Terminated
2008-06-11
Distribution
Nationwide Distribution
Recall number
Z-0688-2008
Recall status
Terminated
CDRH record ID
66461
FDA classified
2008-02-08
Recall event number
45910