Skip to content

Public FDA record

Z-0688-2020: Arrow EZ-IO Intraosseous Vascular Access System 15mm Needle+Stabilizer Kit, Product Code 9018P-VC-005 - P

Recall record Z-0688-2020 for Teleflex Medical. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0688-2020
Company in FDA record
Teleflex Medical
FDA record creation date
Dec 16, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The safety cap attached to needles within the needle sets may become dislodged exposing the needle and potentially causing the needle to protrude through the packaging. If this issue is not detected, the immediate risk of exposure to the affected devices is needle stick injury to the clinician or health care professional. In addition, a puncture of the packaging may compromise the sterility of the needle.

Source facts

Action
Urgent Medical Device Correction notification letters dated 10/8/19 were sent to customers. Customer Action: Our records indicate you have received product that is subject to this correction. Place a copy of this notice with the product to ensure all users are aware of the need to perform this inspection. Removing the kits from the shelf box and placing them into bags (e.g. backpacks) could increase the potential for caps to become loose and product should be visually inspected again at point of use. Our records indicate that you have received products that are subject to this action. We are notifying our customers to take the following actions: 1. Immediately discontinue use and quarantine any products with the catalog number and lot number listed above. 2. Inspect affected products within your control to identify if the safety cap covers the needle, per Figure 1 below, this product is acceptable for use. a) If, after inspection, you identify codes/lots as defective, please: " dispose of such product locally; and " inform us of the affected codes/lots by faxing such information to 1-855-419-8507 or email to recalls@teleflex.com in order for your account to be credited. 3. If you have defective product, please complete the enclosed Acknowledgement Form and fax to [distributor fax number]. 4. If you have no affected stock, please complete the enclosed Acknowledgment Form and fax it to [distributor fax number]. This will allow us to document your receipt of this letter.
Initiated
2019-10-08
Root cause
Package design/selection
Distribution
US nationwide distribution.
Recall number
Z-0688-2020
Recall status
Open, Classified
CDRH record ID
177073
FDA record created
2019-12-16
Recall event number
84054

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗