Skip to content

Public FDA record

Z-0689-2012: MAQUET QUADROX iD PEDIATRIC -OXYGENATOR: BEQ-HMOD 3000; Made In Germany The diffusion membrane oxygenato

Recall record Z-0689-2012 for Maquet Cardiovascular Us Sales, Llc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0689-2012
Company in FDA record
Maquet Cardiovascular Us Sales, Llc
FDA classification date
Jan 11, 2012
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

There are reports of leakage in the Leuer connector of the blood outflow connector of the Quadrox-iD pediatric oygenator.

Source facts

Action
The firm, Maquet Getinge Group, sent an "PRODUCT RECALL URGENT-MEDICAL DEVICE REMOVAL ACTION" letter dated September 22, 2011 to its customers. The letter described the product, problem and actions to be taken. The customers were instructed tocheck their remaining inventory; remove product immediately to prevent use; remove the boxes or individually pouched oxygenators in a secure area of their facility and notify their MAQUET representative (a representative will physically remove the recalled product and replenish the removed units); and complete and return the enclosed Product Removal Disposition Form via fax to: (973)-807-9210 or email: whitney.torning@maquet.com. Should you have any questions or concerns, please do not hesitate to contact MAQUET Customer Service at 800-777-4222.
Initiated
2011-09-22
Root cause
Device Design
Terminated
2014-09-18
Distribution
Nationwide distribution: AL, CA, DE, DC, FL, GA, HI, IL, IN, LA, MI, MO, MS, NC, NE, NJ, NM, NY, OH, OK, PA, and TX.
Recall number
Z-0689-2012
Recall status
Terminated
CDRH record ID
106185
FDA classified
2012-01-11
Recall event number
60673

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗