Open original FDA record ↗Recall record
- FDA identifier
- Z-0689-2016
- Company in FDA record
- Cook Medical Incorporated
- FDA classification date
- Jan 22, 2016
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Manufacturing of the joint between the inner catheter and stylet of the delivery system does not consistently meet specifications, and the joint may separate during deployment. Failure of this joint during a procedure may result in a partial stent deployment. The clinical effect of a partial deployment may be a need for surgical intervention, or may result in vessel trauma or an embolic event
Source facts
- Action
- Cook Medical sent an "URGENT: MEDICAL DEVICE RECALL" letter dated December 22, 2015, to all affected customers. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. Should you have any questions or concerns, please do not hesitate to contact Cook Medical Customer Relations at 1-800-457-4500 or 1-812-339-2235. On 2/5/2016 a second recall notification was sent to affected consignees, expanding the recall to All Lots of the affected product.
- Initiated
- 2015-12-22
- Root cause
- Device Design
- Terminated
- 2017-07-25
- Distribution
- Nationwide Worldwide
- Recall number
- Z-0689-2016
- Recall status
- Terminated
- CDRH record ID
- 142598
- FDA classified
- 2016-01-22
- Recall event number
- 72965