Z-0690-2012: Ultraview SL Command Module is the Spacelabs Multi-parameter Module. It is intended for use with the Pat

Recall record Z-0690-2012 for Spacelabs Healthcare, Llc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0690-2012
Company in FDA record
Spacelabs Healthcare, Llc
FDA classification date
Jan 11, 2012
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

NIBP pump inside the Ultraview SL Command Module, Model 91496, became loose. The pump is suspended by the air hoses and electrical connections, but there can be enough movement to damage internal components and cause the module to fail.

Source facts

Action
SpaceLabs Healthcare sent a Urgent Field Safety Notice dated November 28, 2011, return receipt requested to all US customers. The letter identified the product the problem and the action needed to be taken by customer. On December 6, 2011, SpaceLabs will send a customer letter (translated as necessary) will be emailed to all international subsidiaries and distributors of record. Records of notification will be maintained to ensure complete notification. Customers were instructed If you hear an unusual rattling noise, tag the module, set it aside and contact your Spacelabs field service engineer. Your field service engineer will be contacting you early in 2012 to schedule a mutually agreeable time to upgrade all of your affected modules. Upgrades will be performed on-site and at no cost to you. For any questions about this corrective action program, please call Spacelabs Medical at 1-800-522-7025 and select 2 for Technical Support or, outside the United States, 1-425-657-7200 x5089.
Initiated
2011-11-03
Root cause
Process change control
Terminated
2012-09-19
Distribution
Worldwide Distribution -- USA (nationwide)
Recall number
Z-0690-2012
Recall status
Terminated
CDRH record ID
105892
FDA classified
2012-01-11
Recall event number
60593

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗