Skip to content

Public FDA record

Z-0691-2012: 4D Integrated Treatment Console (4DITC) TrueBeam Linear Accelerator Varian Medical Systems, Palo Alto, CA

Recall record Z-0691-2012 for Varian Medical Systems, Inc. Oncology Systems. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0691-2012
Company in FDA record
Varian Medical Systems, Inc. Oncology Systems
FDA classification date
Jan 11, 2012
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Varian has discovered that on 4D Integrated Treatment Console and TrueBeam, the ad-hoc CIAO(Complete Irradiated Area Outline) calculation results in the CIAO generation that is different from the CIAO in RT Chart. The 4DITC algorithm adds dose for the control point at field edge, which should not be added according to the system specifications.

Source facts

Action
Varian Medical Systems sent a Urgent Medical Device Notice dated December 22, 2011, to all affected customers via certified mail. The letter identified the product the problem and the action needed to be taken by the customer. The Letter will also be distributed to the Varian Sales, Marketing, and Service organizations, informing them of the recall. ***12-20-12, AMENDED INFO ***: An additional letter was sent to consignees on 11/5/12 that Varian identified as also being affected by the recall.*** Canadian account letters were sent by Fed Ex. International Customer Support Managers were sent letters for translation (when necessary) prior to sending to accounts. All of the information will be posted on the Varian web site and all appropriate personnel in the radiotherapy department should be advised of the content of this letter. A complete list of affected code numbers is included as well as contact information for Varian Oncology Help Desk. USA and Canada - 888-827-4265.
Initiated
2011-12-12
Root cause
Device Design
Terminated
2012-07-26
Distribution
Worldwide Distribution -- USA (nationwide)
Recall number
Z-0691-2012
Recall status
Terminated
CDRH record ID
106591
FDA classified
2012-01-11
Recall event number
60829

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 21, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 21, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗