Varian has discovered that on 4D Integrated Treatment Console and TrueBeam, the ad-hoc CIAO(Complete Irradiated Area Outline) calculation results in the CIAO generation that is different from the CIAO in RT Chart. The 4DITC algorithm adds dose for the control point at field edge, which should not be added according to the system specifications.
Source facts
Action
Varian Medical Systems sent a Urgent Medical Device Notice dated December 22, 2011, to all affected customers via certified mail. The letter identified the product the problem and the action needed to be taken by the customer. The Letter will also be distributed to the Varian Sales, Marketing, and Service organizations, informing them of the recall.
***12-20-12, AMENDED INFO ***: An additional letter was sent to consignees on 11/5/12 that Varian identified as also being affected by the recall.***
Canadian account letters were sent by Fed Ex. International Customer Support Managers were sent letters for translation (when necessary) prior to sending to accounts.
All of the information will be posted on the Varian web site and all appropriate personnel in the radiotherapy department should be advised of the content of
this letter. A complete list of affected code numbers is included as well as contact information for Varian Oncology Help Desk. USA and Canada - 888-827-4265.
Initiated
2011-12-12
Root cause
Device Design
Terminated
2012-07-26
Distribution
Worldwide Distribution -- USA (nationwide)
Recall number
Z-0691-2012
Recall status
Terminated
CDRH record ID
106591
FDA classified
2012-01-11
Recall event number
60829